Olympus BX3M-CBFM Manuale di istruzioni - Pagina 23

Sfoglia online o scarica il pdf Manuale di istruzioni per Microscope Olympus BX3M-CBFM. Olympus BX3M-CBFM 44. Control box/fm control box

: Turns ON.
: Blinks.
:Turns OFF.
Phenomena
Number
CB
of beeps
b) When the button of the hand switch for motorized nosepiece (BX3M-HSRE) is pressed, the motorized nosepiece rotates,
but it does not reach the specified hole.
5 times
: Varies depending on state.
Cause
HS
MIX
The setting of the
dip switch is not
correct.
The motorized
nosepiece did not
reach the click
position within the
specified time.
The motorized
nosepiece is
damaged.
BX3M-CB/CBFM
Remedy
Set the main switch to
(OFF),
and use the dip switch on the side
of the control box (BX3M-CBFM
or BX3M-CB) to set the number of
objective mounting holes on the
nosepiece.
Set the main switch to
(OFF)
and check that there is no obstacle
when the nosepiece rotates.
If the nosepiece contacts with the
obstacle, remove the obstacle.
Then, set the main switch to (ON).
If the same phenomena still occurs,
contact Olympus.
Page
9, 33
9
19