Olympus BX41-P Manuale di istruzioni - Pagina 23

Sfoglia online o scarica il pdf Manuale di istruzioni per Microscope Olympus BX41-P. Olympus BX41-P 36. Polarizing microscope
Anche per Olympus BX41-P: Manuale di istruzioni (36 pagine)

Olympus BX41-P Manuale di istruzioni
Conoscopic image
Fig. 26
²
Fig. 27
6. To brighten and facilitate observation of the conoscopic image during
optical axis adjustment, rotate the analyzer slightly away from the position
of total extinction. (Fig. 26)
7. Insert the two provided centering wrenches ² into the centering holes
for the primary hole with the 20X or 10X objective on the nosepiece.
8. Manipulate the centering wrenches ² to bring the bright center portion
of the conoscopic image into the center of the field of view (Fig. 27)
9. Pull out the Bertrand lens knob to the OUT (¦) position. At this state,
center the condenser again, as in step 4 above.
BX51/52-P,BX41-P
20