Olympus BX41-P Manuale di istruzioni - Pagina 26
Sfoglia online o scarica il pdf Manuale di istruzioni per Microscope Olympus BX41-P. Olympus BX41-P 36. Polarizing microscope
Anche per Olympus BX41-P: Manuale di istruzioni (36 pagine)
Fig. 31
23
3
Adjusting the Eyepiece Cross Lines
}To align the eyepiece cross lines and the vibration direction, the
orientation plate provided with the U-CPA or U-OPA is required.
Orientation plate (U-PJ: Single module name)
# Remove the test plate adapter and the compensator from the light
path.
1. Swing out the top lens of the condenser.
2. Place the orientation plate on the stage. Remove the analyzer from the
light path, and use the 4X objective to focus.
(Fig. 31)