Mounting the Objectives
# Insert the 10X or 20X objective into the primary hole (position where
black rubber plugs are inserted into the centering holes on the nose-
piece).
Mount the other objectives in such a manner, that the magnification in-
creases from low to high power in a clock-wise direction from the primary
hole.
4
Mounting the Revolving Nosepiece (U-P4RE)
1. Turn the coarse adjustment knob to lower the stage all the way.
2. Using the Allen screwdriver, loosen the nosepiece clamping screw @ on
the microscope frame.
3. Carefully slide the nosepiece along the dovetail to its limit, in the direction
of the arrow.
4. Clamp the nosepiece by tightening the nosepiece clamping screw.
BX51/52-P
(Fig. 4)
(Fig. 5)