5-5 Objectives

Fig. 32
1
Setting the Correction Collar
A culture microscope is designed to observe specimens contained in
vessels of various bottom thicknesses. In order to achieve optimum
objective performance of the culture microscope, the high N . A . objec-
tives are provided with a correction collar @. (Available on the
LWDCDPlan40XPL-6 only.)
@
The correction collar is effective with a vessel bottom from 0 to 2 mm
thickness.
1. If the thickness of the vessel bottom is known:
Match the correction collar to the thickness of the vessel bottom using
the collar scale provided.
2. If the thickness of the vessel bottom is unknown:
The optimum position for the correction collar can be obtained from the
image resolution. If a satisfactorily sharp image is not obtained after fo-
cus adjustment, rotate the correction collar to the right and left so that
you can compare the images at both sides. Reset the collar to the better
image, then starting from this position, further rotate the collar to the right
and left until both images can be obtained for comparison. By repeating
this procedure several times, you will find best position for the correction
collar. Refocus after rotating the correction collar.
CK30/CK40
(Fig. 32)
18