1 Preparation for Operation
1-1 Turning the Power On
Ar laser power supply (Type 300)
HeNe Green/Red laser power supply
Preparation for Operation
1. Set the power switches of the following units to I (ON).
Power Supply Unit FV10-PSU
Mercury Burner Power Supply Unit
Microscope Control Box BX-UCB or IX2-UCB
2. Set the power switches of the PC and monitor to I (ON).
3. Turn on the lasers as follows.
3.1 Argon laser: FV5-LA-MAR
Set the power switch to ON. (This starts the fan of the laser.)
Turn the key to the ON position.
It takes a few tens of seconds after the key is set to ON till the
laser oscillation begins.
3.2 Helium-Neon Green/Red Laser: FV5-LA-HEG/HER
Turn the key to the I (ON) position.
It takes a few tens of seconds after the key is set to ON till the
laser oscillation begins.
NOTE
To ensure stable laser light output, it is recommended to
warm up the laser power supply after turning it on. The
warm-up period should be 10 minutes or more when using
the Argon laser power supply and 30 minutes or more when
using the Helium-Neon Green or Red laser power supply.
/ Turning the Power On
II. PREPARATION For OPERATION
II. 1 - 1
Page