Olympus U-25ND25 Manuale di istruzioni - Pagina 16
Sfoglia online o scarica il pdf Manuale di istruzioni per Microscope Olympus U-25ND25. Olympus U-25ND25 42.
@ ² ³
Fig. 3
A
B
C
D
7
Centering the Mercury Burner
|
}Set the main switch to " I " (ON) and wait for 5 to 10 minutes until the arc
image stabilizes before proceeding to the mercury burner centering.
1. Engage the shutter in the light path by pressing the RSHT button on the
hand switch.
2. Engage the B or IB excitation fluorescence mirror unit in the light path (If
these mirror units are not available, engage another fluorescence mirror
unit in the light path. Also note that, when using an U excitation fluores-
cence mirror unit, be always sure to observe the specimen image through
a UV cut plate.)
3. Engage the 10X objective in the light path by pressing the objective
button, place the U-CST centering target on the stage, and adjust the
centering of the center of crosslines on white surface of the U-CST with
respect to the center of field of view.
4. Engage the empty place (the objective cap should be removed) in the
light path.
5. Pull out the aperture iris diaphragm knob @ (to minimize it) and push in
the field iris diaphragm knob ² (to maximize it).
6. Release the shutter by pressing the RSHT button.
7. Project the arc image on the U-CST by turning the collector lens focusing
knob ³. (A)
If the arc image is not projected, adjust the burner centering knobs |.
8. Bring the arc image on the center of the left (or right) half of the field by
turning the burner centering knobs |. (B)
9. Fit the Allen screwdriver in the mirror focusing screw (ƒ in Fig. 4) on the
rear of the lamp housing and adjust to bring the mirror arc image in
focus. (C)
10.
Overlay the arc image with the mirror arc image by turning the burner
centering knobs |. (D)
}During observation, adjust the collector lens focusing knob ³ so that the
observed field is uniform.
}Hereafter, the mercury burner centering need not be adjusted until the
next time the mercury burner is replaced.
(Fig. 3)
13