8
ASSEMBLY
8-1 Assembly Diagram
The diagram below shows the sequence of assembly of the various modules. The numbers indicate the order of
assembly.
The module numbers shown in the following diagram are merely the typical examples. For the modules with which the
module numbers are not given, please consult your Olympus representative or the catalogues.
When assembling the microscope, make sure that all parts are free of dust and dirt, and avoid scratching
CAUTION
any parts or touching glass surfaces.
Assembly steps enclosed in
} All assembly operations are possible by using the Allen screwdriver (
The Allen wrench (
the performance, have your Olympus representative conduct this work.)
· Be sure to insert the sliders in the orientations shown in the diagram. Otherwise, they cannot be fitted in click
CAUTION
positions and engaged correctly in the light path.
· To prevent fire, install the lamp housing with the heat radiating fins positioned on the top and with sufficient
space around the housing.
Eyepiece
Fluorescence
mirror unit
Turret cover
5
Anti glare plate
Objectives/Revolving
nosepiece
will be detailed on the subsequent pages.
) provided with the illuminator is used only for clamping the screws inside the illuminator. (to retain
Observation
tube
Illuminator
clamping
screws
2
1
6
Light shield sheet
Heat radiating fins
Mercury burner
USH-103OL
3
Lamp housing
clamping screws
4
Illuminator
BX3-URA
BX3-RFAS
Microscope frame
BX3 series
) provided with the microscope.
Lamp housing for
mercury burner
U-LH100HG
U-LH100HGAPO
Power supply unit
U-RFL-T
ND filter slider
(BX3-RFAS)
ND filter (BX3-URA)
BX3-25ND6
BX3-25ND25
ND filter (BX3-URA)
BX3-25ND6
BX3-25ND25
Analyzer
U-AN-2
24