Chapter 3 Preparation and Inspection
14
 If Power Fails to Come on
If power fails to come on, inspect the system as follows:
1.
Confirm that the Power Cord is connected securely to a Wall Mains Outlet
and the Power Cord Receptacle on the Maintenance Unit.
2.
Confirm that the Circuit Breakers have not tripped by checking that the
breaker buttons have not popped up. If the buttons have popped up, reset
the breakers according to the following procedure 3. and 4.
Reset the Circuit Breakers only once. If the power is still not
supplied to the Maintenance Unit after resetting the Circuit
Breakers, immediately turn the Power Switch OFF "
disconnect the Power Cord from a Wall Mains Outlet and
contact Olympus. This condition can result in electrical shock
and/or fire.
When the Circuit Breakers trips, do not reset the Circuit
Breakers immediately. Turn the Power Switch OFF "
at least 15 seconds before resetting the Circuit Breakers to
prevent Circuit Breaker damage.
3.
Turn the Power Switch OFF "
the Circuit Breaker buttons until it stops with a click.
4.
Reconnect the Power Cord and turn the Power Switch ON "
that power is being supplied to the Maintenance Unit.
MAINTENANCE UNIT MU-1
", disconnect the Power Cord and depress
"
". Wait
". Confirm