AccuLine 40-6540 Manuale di servizio - Pagina 8
Sfoglia online o scarica il pdf Manuale di servizio per Strumenti di misura AccuLine 40-6540. AccuLine 40-6540 13. Automatic rotating laser
4.1.2. Characterize the type of error
Use the following rules to determine what type of accuracy error the instrument for test has, and reference the
stated section to adjust for it. Note that the unit my have a combination of different types of errors to adjust for, in
which case, multiple calibration must be performed.
Accuracy is beyond
published specification
A=0 & B=0
1
No
A and B <> 0 but
A + B = 0
A and B <> 0 but
A = B
A and B <> 0 and
A + B <> 0
No
Prepared By: Tim Wojo
Yes
Section 4.1.2.1
Yes
Oblique Error
No
Yes
Section 4.1.2.2
No
Yes
Section 4.1.2.1
Section 4.1.2.2
No
1
Revision - Date: 0 – 20080430
Yes
Yes
Taper/Cone
Error
Yes
Oblique Error
Taper/Cone
Error
Yes
A = B <>C=D
40-6540 Service Manual
Accuracy meets published
specification
Yes
C=0 & D=0
No
C and D <> 0 but
C + D = 0
No
C and D <> 0 but
C = D
No
C and D <> 0 and
C + D <> 0
No
Section 4.1.2.3
Wave Error
Page 8 of 13