Olympus CIX100 Manuale di configurazione dei sistemi - Pagina 25

Sfoglia online o scarica il pdf Manuale di configurazione dei sistemi per Apparecchiature di laboratorio Olympus CIX100. Olympus CIX100 38. Cleanliness inspector system
Anche per Olympus CIX100: Manuale di installazione (14 pagine), Manuale di istruzioni (46 pagine)

Olympus CIX100 Manuale di configurazione dei sistemi
9

Care and maintenance

Your OLYMPUS CIX100 Cleanliness Inspector System doesn't require any
maintenance. If you have any questions or problems, please contact our
customer service:
• E-mail: [email protected]
• Phone: (+ 49) 251-79800-6444, fax: (+ 49) 251-79800-6445
9.1

Cleaning the particle standard device

Make sure to keep the particle standard device clean and free of dust. Don't
touch the particle standard device with your fingers, to avoid getting grease
on it. If necessary, clean the particle standard device with a rubber dust
blower. For more persistent marks, we recommend cleaning the particle
standard device with a lint free precision wipe and some acetone or alcohol.
Follow the manufacturer's safety instructions for the substance that you are
using.
10

Troubleshooting

If questions or problems come up that you can't solve with the help of these
instructions, Olympus Soft Imaging Solutions Customer Service will be
happy to help you:
• E-mail: [email protected]
• Phone: (+ 49) 251-79800-6444, fax: (+ 49) 251-79800-6445
Problem
The recorded images are not correctly aligned with each other.
If you notice when checking the system or during the inspection of the
sample that the recorded images are not assembled correctly, it may be that
the calibrations are no longer valid.
OLYMPUS Cleanliness Inspector CIX100
When the images have been assembled this
way, it indicates that the
Magnification
Calibration] calibration is no
longer valid.
25
9 Care and maintenance
[Manual
System Setup Guide