Olympus CIX100 Manuale di istruzioni - Pagina 24

Sfoglia online o scarica il pdf Manuale di istruzioni per Apparecchiature di laboratorio Olympus CIX100. Olympus CIX100 46. Cleanliness inspector system
Anche per Olympus CIX100: Manuale di configurazione dei sistemi (38 pagine), Manuale di installazione (14 pagine)

Olympus CIX100 Manuale di istruzioni
OLYMPUS CIX100 Cleanliness Inspector System
6

Calibrating White Balance for a new sample type

Make sure to log into the CIX ASW software as administrator. Then you have full
access to the [System configuration] on the start page of your CIX ASW software.
Calibrate the stage limits as described in the CIX ASW software's help. Insert the
new sample type with the new sample, which you activated in the [Hardware] >
[Device List] > [Device settings] dialog box as described in chapter "Adding new
sample types to your CIX100 system" on page 13. These may be
tape lift,
l
particle trap, or
l
covered membrane.
l
For a list of samples and applications, please refer to chapter "Available sample
holders and sample types" on page 5 of this manual.
1. Click the [Calibration] button on the start page of CIX ASW.
24