Olympus CKX41 Manuale di riparazione - Pagina 19
Sfoglia online o scarica il pdf Manuale di riparazione per Apparecchiature di laboratorio Olympus CKX41. Olympus CKX41 43. Reflected fluorescence system
Anche per Olympus CKX41: Panoramica (7 pagine), Opuscolo e specifiche (4 pagine), Manuale di istruzioni (36 pagine)
CKX31/CKX41
No.
Parts name
(A)
COVER
(B)
COVER
(C)
COVER
(D)
BINOCULAR TUBE CUKK3X6SA, 4pcs. (*4)
(E)
INDICATOR PLATE
(F)
DIOPTER RING
(G)
HELICOID ASS'Y
(H)
SLEEVE ASS'Y
(I)
PRISM ASS'Y
(J)
PRISM ASS'Y
(K)
TORQUE PLATE
(L)
SPRING WASHER
(M)
WASHER
(N)
MOUNT
(O)
RING
(P)
PRISM MOUNT-R
(Q)
WASHER
(R)
PRISM MOUNT-L
(S)
RING
(T)
SHAFT
(U)
FIXING PLATE
(V)
PRISM
* In this manual, left ( right) sleeve indicates the left (right) side with the observed condition.
C. DISASSEMBLY AND ASSEMBLY PROCEDURES
Screw
3PUTB2X4SB, 3pcs. (*1)
3PUTB2X4SB, 3pcs. (*2)
CUTB3X10SA, 2pcs. (*3)
ACU3X3SA, 2pc. (*5)
HU3X4SA, 1pc. (*6)
CUK3X6SB, 2pcs. (*7)
CUK3X6SB, 2pcs. (*8)
CUK3X8SA, 2pcs. (*9)
CUK3X8SA, 2pcs. (*10)
AWU3X4SA, 2pcs. (*11)
CUK3X6SA, 2pcs. (*12)
C-8
Grease Adhesive
Attach it with double coated
adhesive tape.
Since the mounting procedure is
the samet as that of CKX31,
refer to C-6.
Refer to D-2.
Left/right optical axis:
Follow the adjustment procedure
on D-8.
Revolving axis:
Follow the adjustment procedure
on D-8.
Push it in the arrow directions
during assembling.
Exit pupil center:
Follow the adjustment procedure
on D-8.
Same as the above
Tension: standard 5-15N
Since the adjustment is the same
as that of CKX31, refer to C-6.
OT3189
OT3189
OT3189
OT3189
Apply grease to the contacting
surface only.
Absolute optical axis:
Follow the adjustment procedure
on D-8.
Remark
(OT3223)