Olympus CKX41 Manuale di riparazione - Pagina 37

Sfoglia online o scarica il pdf Manuale di riparazione per Apparecchiature di laboratorio Olympus CKX41. Olympus CKX41 43. Reflected fluorescence system
Anche per Olympus CKX41: Panoramica (7 pagine), Opuscolo e specifiche (4 pagine), Manuale di istruzioni (36 pagine)

Olympus CKX41 Manuale di riparazione
CKX31/CKX41
<Optical adjustment in binocular tube>
Adjustment item
1. Revolving axis
(right sleeve)
2. Left/right optical axis
(left sleeve)
3. Absolute optical axis
(right sleeve)
4. Exit pupil center
5. Parfocality
* Parfocality: Check the parfocality in right sleeve. If not meet the standard, confirm that the screws fixing the
sleeve is not loosened. (In case where the screw is loosened, adjust the parfocality by moving
the sleeve vertically because the sleeve is possible to be displaced.)
D. REPAIR PROCEDURE
FT-36
KN0048 (with adapter-1)
KN0029
KN0003
Eyepiece(Jig)
1) Standard eyepiece
(KN0048; Adapter-1)
1) Standard eyepiece
(KN0048; Adapter-1)
1) Standard eyepiece
(KN0048; Adapter-1)
1) Centering telescope
(KN0029)
1) Standard eyepiece
(KN0048; Adapter-1)
2) Focusing telescope
(FT-36 or U-FT)
D-10
KN0041
PlanC10X
Objective(Jig)
2) PlanC10X
2) PlanC10X
2) Standard objective
(KN0041)
2) PlanC10X
3) Standard objective
(KN0041)
Jigs and tools
3) Test plate :
concentric circles
(KN0003) or
cross micrometer
(OB-M1/100SQ)
3) Test plate :
concentric circles
(KN0003) or
cross micrometer
(OB-M1/100SQ)