Olympus CKX41 Manuale di istruzioni - Pagina 23

Sfoglia online o scarica il pdf Manuale di istruzioni per Apparecchiature di laboratorio Olympus CKX41. Olympus CKX41 36. Reflected fluorescence system
Anche per Olympus CKX41: Panoramica (7 pagine), Opuscolo e specifiche (4 pagine), Manuale di riparazione (43 pagine)

Olympus CKX41 Manuale di istruzioni

SPECIFICATIONS

Module
Reflected Fluorescence Illumi-
nator
CKX-RFA
Lamp Housing for 50W Mercury
Burner
U-LH50HG
Power Supply Unit
U-RFLT50-100
U-RFLT50-200
Operating environment
Fluorescent lighting method: Reflected fluorescent lighting based on slide
Fiend number: 22 (when the WH10X is used).
Shutter: Switching using filter slider (Shutter/Open position/Filter pocket ).
Excitation light protective shield: Fixed on condenser with a screw.
Possible observation methods:B-excitation, G-excitation, U-excitation
Burner centering, Up/down, left/right.
Mirror centering: Up/down, left/right.
Burner focusing: Collector focusing and mirror focusing screw system.
Safety mechanism: Interlock switch, UV cut filter built in.
<< Applicable high-pressure mercury burners >>
· HBO50W/AC (OSRAM)
· CS50W4 (PHILIPS)
Ignition system: Auto ignition.
Hour counter: Displays the accumulated hours of operation.
Input rating: Input voltage switchable.
U-RFLT50-100: 100/110/120 V AC
U-RFLT50-200: 220/230/240 V AC
Dimensions
U-RFLT50-100: 150(W) x 150(H) x 320(D) mm
U-RFLT50-200: 150(W) x 150(H) x 210(D) mm
Weight
U-RFLT50-100: 12 kg (26.4 lbs)
U-RFLT50-200: 7 kg (15.4 lbs.)
Indoor use.
Altitude: Max. 2000 meters.
Ambient temperature: 5° to 40°C (41° to 104°F).
Maximum relative humidity: 80% for temperatures up to 31°C (88°F ),
Supply voltage fluctuations; Not to exceed ±10% of the normal voltage.
Pollution degree: 2 (in accordance with IEC60664).
Installation category (Overvoltage category): II (in accordance with IEC60664).
REFLECTED FLUORESCENCE SYSTEM
Specifications
switching of fluorescent mirrors.
(optional).
, 2.5/2.3/2.0 A, 50/60 Hz
, 1.7 A, 50/60 Hz
decreasing linearly through 70% at 34°C
(93°F ), 60% at 37°C (99°F), to 50% relative
humidity at 40°C (104°F).
FOR CKX41
20