Olympus SZX-RFL3 Manuale di istruzioni - Pagina 4

Sfoglia online o scarica il pdf Manuale di istruzioni per Apparecchiature di laboratorio Olympus SZX-RFL3. Olympus SZX-RFL3 22. Fluorescence attachment

1
Getting Ready
1. This manual pertains only to the fluorescence attachment. Before using this attachment together with the SZX micro-
scope and associated options, make sure that you have carefully read and understood their manuals, and understand
how the microscopic system should be used together.
2. The fluorescence attachment is a precision instrument. Handle it with care and avoid subjecting it to sudden or severe
impact.
3. Do not use the equipment where it is subjected to direct sunlight, high temperature and humidity, dust or vibrations. (The
operating environment should be identical to that of the SZX microscope body.)
4. Always use the power cord provided by Olympus. Before plugging the power cord, make sure that the main switch of the
power supply unit is set to "
5. To allow heat from the unit to dissipate well, reserve a distance of at least 10 cm between the lamp housing and power
supply unit.
6. The power cord can be used to cut the power supply in case of emergency. To make this possible, the power supply unit
should be installed so that the power input connector (on the rear of the power supply unit) or the power outlet is easily
accessible for unplugging in case of emergency.
Maintenance and Storage
2
1. Clean all glass components by wiping gently with gauze. To remove fingerprints or oil smudges, wipe with gauze slightly
moistened with a mixture of ether (70%) and alcohol (30%) .
Since solvents such as ether and alcohol are highly flammable, they must be handled carefully. Be sure to
keep these chemicals away from open flames or potential sources of electrical sparks –– for example,
electrical equipment that is being switched on or off. Also remember to always use these chemicals only in
a well-ventilated room.
2. If any part of the equipment (other than glass components) gets dirty, wipe it with a clean cloth. If the part is extremely dirty,
do not use organic solvents to clean it; instead, use a soft, lint-free cloth lightly moistened with a diluted neutral detergent.
3. Do not disassemble any part of the equipment. This could result in malfunctions or reduced performance.
4. The burner has a service life period of 200 or 300 hours. When the hour counter on the power supply unit indicates 200
hours (with the USH102D) or 300 hours (with the HBO103W/2), replace the burner with a new one. (Page 11)
Caution
3
If the equipment is used in a manner not specified by this manual, the safety of the user may be imperiled. In addition, the
equipment may also be damaged. Always use the equipment as outlined in the instruction manual.
The following symbols are used to set off text in this instruction manual.
: Indicates that failure to follow the instructions in the warning could result in bodily harm to the
user and/or damage to equipment (including objects in the vicinity of the equipment).
# : Indicates that failure to follow the instructions could result in damage to the equipment.
} : Indicates commentary (for ease of operation and maintenance).
" (OFF).
SZX–RFL3
2