5.3. Inspection
1. Inspect all components of the device
for visible soil under good lighting
conditions. Inspect the integrity of
O-rings.
2. Inspect linen for obvious tears and rips in
fabric or seams and check for integrity of
the buckles. Invert the linen and inspect
integrity of seam tape.
TEAR IN FABRIC: If there is a tear, contact
Arbutus Medical for replacement information.
Inspection with a bright, cool light inside of
the Cover is necessary to detect small tears.
STAINS: Check for blood/other stains. If
found, send for re-cleaning. (Note: Dye residue
is acceptable).
TAPE or RESIDUE: Remove any tape and
residue. If you cannot remove, send for re-
cleaning.
3. Invert the linen back to the right side and
place it over a light fixture with bright
cold light source (eg. fluorescent light
bulb). Look for pin holes in fabric. Move
the linen around on the light fixture to
ensure all areas are inspected. Preferably,
do this step in a dark room to make it
easier to spot pin holes.
4. If linen passes inspection, record each use
on the tracking grid located on the inside
of the linen, by marking one square using
a laundry pen. If linen fails inspection,
remove it from clinical use.
DrillCover Hex System: Instructions For Use
15