Accurex DSX Series Manuale d'uso - Pagina 7
Sfoglia online o scarica il pdf Manuale d'uso per Bilance Accurex DSX Series. Accurex DSX Series 10.
(1)
Put enough samples on the scale, press【TARE】key to complete the sampling.
(2) By pressing the 【UNIT】 key, the 'WEIGHT' window displays '000000',press 【 ↑ 】 key to adjust
the number of weights, press【TARE】 key to confirm.
When the weight of the sample is less than 0.1% of the maximum capacity of the scale, the 'WEIGHT'
window displays '-CSL-', which means the sample is insufficient and needs to be sampled again.
Press 【 SETUP 】 key to re-sample, press 【 MODE 】 key to return to the weighing mode.
When percentage sampling is complete, the 'WEIGHT' window displays the current weight as a
percentage of the sample weight, accurate to 0.1%.
Under sampling mode, press 【 SETUP 】 key to exit sampling mode.
4. [Set Alarm] :
Under the weighing mode or counting mode, long press 【 UNIT 】 key, the 'WEIGHT' window displays
'HL---ZL'. Press 【 ↑ 】 key to switch between 'HL---ZL' and 'HL---PCS'. 'HL---ZL' stands for set weight
alarm, while 'HL---PCS' stands for set quantity alarm, and press【TARE】key to confirm.
The 'WEIGHT' window display with the last set alarm upper limit and the prompt '-HH-' in turn, if you
do not need to change the previous set value, press 【TARE】 key to directly enter the next step to
set the alarm lower limit; if you need to change the previous set value, press 【 ↑ 】 key to input the
number, press 【TARE】key to confirm current value, and began to set the next, if it is the most at
the end of the value, then automatically enter the next step to set the alarm lower limit.
If you just enter the state of setting alarm lower limit, the 'WEIGHT' window displays the last set
weight lower limit and prompt '-LL-' in turn. The operation of this process is the same as setting alarm
upper limit. After the setting of alarm lower limit, enter the setting of alarm mode.
When setting the alarm mode, the 'WEIGHT' window displays '-IN-' stands for set alarm within this
range, '-OUT-' stands for set alarm without this range, and '-NO-' stands for no alarm. At this time,
press 【 ↑ 】 key to switch and press【TARE】key to confirm.
During the setting alarm process, press 【 SETUP 】 key to exit the setting, and the current setting
value is invalid.
5. [Weight Calibration] :
After starting up the scales, long press 【 SETUP 】 key to enter the calibration.
The 'WEIGHT' window displays 'D-----5', press 【 ↑ 】 key to adjust the division value within'1/2/5/10/20',
and press 【TARE】 key to confirm.