DigiFLO HH Manuale - Pagina 4

Sfoglia online o scarica il pdf Manuale per Strumenti di misura DigiFLO HH. DigiFLO HH 5. Concentrator analyzer

Specification:
Size
Weight
with battery
Response Time
Power On Duration
Accuracy
Range
Max pressure
Temperature range
Power consumption
Battery
Battery life
Limited Warranty
Blinking Alarms
BATTERY
Reasons to replace the battery:
 BAt is seen blinking on the display.
 DigiFLO Concentrator Analyzer does not respond to a Power On.

CHANGING THE BATTERY

1. Open the rear battery compartment.
Caution: Observe proper battery installation.
2. Carefully lift the battery from its location.
3. Remove the clips from the battery.
4. Apply the clips to a new battery.
5. Position the new battery within its groove, with the clips facing the Inlet.
6. Turn ON the unit. The display should turn ON. If it does, close the battery compartment. If it
does not, replace the battery with another.

TROUBLESHOOTING

There are no user serviceable parts within the DigiFLO Concentrator ANALYZER.
If the unit fails to come ON, replace the battery with a known good battery.
If the unit fails to come on with known good battery, contact DigiFLO Technical Service for repair.
If the unit comes on and the reading are erratic or frozen, contact DigiFLO Technical Service for repair.
If the front button does not respond, contact DigiFLO Technical Service for repair.
If a segment within the display is missing or is displayed intermittently, contact DigiFLO Technical
Service for repair.
9" x 1.5" x 1"
164g
0.1 Second
2 minutes
Concentration: ±1.5%
Flow Rate: ±0.1 LPM
Pressure ±0.5%
●Concentration: 21% - 95.7% (Oxygen Concentrator range)
●Flow Rate: 0 to 20 LPM
●Pressure: vacuum to 241 KPA
280 KPA
-45 to +70°C
10ma
Single Alkaline 9V battery
1000+ read cycles
5 years
OFF – Unit operational time of 2 minutes has expired
BAt – Low battery indicator
4