IEC/EN 61000-4-8
Conducted Immunity
IEC/EN 61000-4-6
Irradiated Conducted
IEC/EN 61000-4-3
Note U
is the value of the power supply voltage
T
Antibacterial Filter: The filter is produced with (PTFE) hydrophobic material witch prevents fluids entering the
pump unit and also offers protection to prevent bacterial contamination of the pump unit. When the filter is wet, it's
not possible to use the unit therefore the filter should be changed immediately.
In case of possible contamination or discoloration of the filter, change the filter immediately.
Do not use the instrument without the filter properly installed.
To clean the plastic housing of the device wear disposable protective gloves and clean with denatured alcohol or
hypochlorite solutions.
Washing and / or cleaning the autoclavable collection bottle can be carried out as follows:
Wear protective gloves and apron (glasses and face mask if necessary) to avoid contact with
contaminating substances;
Disconnect the bottle from the device
Disconnect all tubes from the jar and the filter
Empty and properly dispose of the contents in the collection bottle
Separate all parts of the cover (overflow valve, o-ring);
Thoroughly rinse all parts in cold water then soak in warm water (temperature not to exceed 60°C). Wash
thoroughly and if necessary use a non-abrasive brush to remove and deposits. Rinse in running warm water and
dry all parts with a soft cloth (non-abrasive).
Do not use leave bleach solutions in the collection bottle. If required for disinfection, use a 10%
bleach to water solution for cleaning, but then rinse thoroughly with water. Bleach can damage the
polycarbonate material of the collection bottle.
3Vrms 150kHz to 80MHz
(for appliances that aren't life -
supporting)
3V/m 80MHz to 2.5 GHz
(for appliances that aren't life -
equipment)

ACCESSORIES SUPPLIED

DESCRIPTION
COMPLETE COLLECTION BOTTLE
2000ml
TUBING SET (3 pieces)
HANDHELD VACUUM CONTROLLER
2 ADAPTERS (PIPETTE TIP AND FINE
POINT METAL TUBE)
ANTIBACTERIAL FILTER

CLEANING OF ACCESSORIES