Accuris W3100A-120 Manuale operativo - Pagina 6

Sfoglia online o scarica il pdf Manuale operativo per Bilance Accuris W3100A-120. Accuris W3100A-120 16. Analytical balance w3100 series

Accuris W3100A-120 Manuale operativo

3. OPERATING YOUR BALANCE

NOTE: To avoid dust from entering the internal weighing mechanism,
keep your balance door closed whenever it is not in use.
Plug in and pre-warm your balance at least 30 minutes before use.

3-1 Basic weighing

 Press the power key, the balance will turn on and display 0.0000g
 Place a weigh boat or weighing container on the weighing pan
 Press the TARE key and wait for the display to show 0.0000g
 Place the sample onto the container and close the draft shield doors
 The display will stabilize and the weight will be displayed
 Repeat the steps to weigh the next sample

3-2 Count weighing

This function is used to determine a total number of pieces of similar
weights. A known number of pieces are weighed as a reference, and the
average weight of each piece is automatically calculated. An unknown
quantity of pieces can then be calculated.
1. First set the sample quantity that will be used as the reference
quantity weight. See the section 5-2 for PARAMETER SETTING.
For example, for a reference quantity weight of 50 pcs, set up
function C2 at setting 2 (display will read C2--2.)
2. Press the TARE key, the balance will display 0.0000g.
3. Press the MODE key repeatedly until the units "PCS" is displayed.
4. Place the sample on the pan, then press the SET key, the balance will
read the amount number of the samples.
Press the MODE key to change out of the counting mode.

3-3 Percent Deviation

This mode is used to calculate the percentage of weight a sample varies
6