Accuris W3100A-210-E Manuale operativo - Pagina 7
Sfoglia online o scarica il pdf Manuale operativo per Bilance Accuris W3100A-210-E. Accuris W3100A-210-E 16. Analytical
from a reference weight.
1. Empty the weigh pan;
2. Press the TARE key to zero the display;
3. "0.0000 g" will be displayed; Press MODE key till the display is %
mode.
4. Place the reference weight onto the weigh pan, and close the doors;
5. Press the SET key. 100% will be displayed;
6. Remove the reference weight from the weigh pan;
7. Place an object to be compared to the reference weight onto the
weigh pan and close the draft shield doors.
8. A percentage value of the deviation between reference and the
sample will be displayed.
3-4 DISPLAY AND CONTROL PANEL
DISPLAY SYMBOLS
g
ct
Lb
oz
pcs
%
○
+
-
·······
CAL in
CAL dn
CAL ...
CAL up
CAL-no
CAL end
+E
-E
------
SAVE ---
S --- END
Gram
Carat
Pound
Ounce
piece counting mode
percentage mode
Indicates the weight is stable
positive indicator
negative indicator
Waiting
Calibration starting
Load the appropriate calibration weight
Calibrating
Unload the calibration weight
Not calibrated properly
Calibration complete
over the max capacity
under the readability
Set key valid
.
Reading the data
Saving
Saving complete
7