Safety Information, Mechanical
WARNING
CAUTION
SIREMOBIL Iso-C
Area dose product measurement device
Danger of injuries on mechanical parts!
Non-compliance can lead to slight to medium injuries, especially
of the hands.
The C-arm of the system is manufactured from carbon fiber com-
posite material (CFK). In the case of damage to the carbon fiber
structure, for example caused by collision or by cutting or scrap-
ing tools, individual carbon fibers or carbon fiber bundles can
stand out and on contact cause injuries to the skin or penetrate
into the tissue of the hand.
Before starting work on the C-arm, make sure that there is no
damage to the carbon fiber structures (visual inspection).
If smaller superficial damage is detectable in the painted area or
in the area of the running surface edges of the C-arm, wear work-
ing gloves and remove any projecting carbon fibers. Then care-
fully rub the damaged places flat with fine emery cloth and close
the damaged surface with paint. A spray paint in the color of the
C-arm can be used for this purpose (See Service Tool Catalogue).
Remove any carbon fibers lying in the C-arm profile with a damp
sponge or cloth.
If greater damage to the carbon fiber structure indicating general
or large-scale destruction of the carbon fiber structure (e.g. crack
formation or flaked off places in the running surfaces with torn
carbon fibers) can be detected, then replace the C-arm.
Danger of injuries on mechanical parts!
Non-compliance can lead to slight to medium injuries, especially
of the hands.
After covers are opened, parts such as flat plugs, threaded bolts,
cut-off cable ties and edges of components can be touched
which, if care is not taken, can cause crushing, scrapes and cuts
to the skin, particularly to the hands.
Perform the particular work steps with special care and attention
to detail.
If needed, wear working gloves.