Numerical values on a linear scale of 0 to 8 kilopascals (kPa) are repre­
sented by a color scale ranging from violet to red, when using the
default window/level of the relative stiffness image. The scale defines
estimated tissue stiffness: very soft tissues in the 0-2 kPa range are
shown with a color range of violet to blue, whereas stiffer tissues above
4 kPa are depicted with colors ranging from green through yellow to
red.
The numerical values corresponding to the color scale are obtained by a
mathematical analysis of wave images with a locally homogeneous, lin­
ear elastic model that represents tissue stiffness as directly proportional
to the square of the local wavelength. Since the local wavelength at a
given pixel is calculated from a region surrounding it, the stiffness esti­
mate is most valid when the tissue is uniform for at least 1 cm in all
directions to avoid partial volume effects. The validity of the local stiff­
ness estimate depends on the presence of waves of sufficient quality (as
indicated in the "Relative Stiffness Map 95 %" images). Other factors
that can affect the quality of the stiffness images are patient motion
during the acquisition, failure to place the passive driver in the recom­
mended location for liver imaging, and failure to acquire liver imaging
data during end expiration.
This scale is meant to provide you with a guide to distinguish between
the stiffness variations of the tissue within each acquired image.
Wave Image
Before applying the Helmholtz equation, the phase difference images
are unwrapped. A least squares fit is applied to remove global displace­
ments due to bulk motion, leaving just the shear waves. The unwrapped
and corrected wave displacements are displayed in this series.