Siemens ULTRAMAT 6F Manuale di sicurezza - Pagina 12

Sfoglia online o scarica il pdf Manuale di sicurezza per Apparecchiature industriali Siemens ULTRAMAT 6F. Siemens ULTRAMAT 6F 24. Gas analyzer for the determination of ir-absorbing gases
Anche per Siemens ULTRAMAT 6F: Manuale dell'attrezzatura aggiuntiva (20 pagine), Manuale di istruzioni per l'uso (49 pagine)

Siemens ULTRAMAT 6F Manuale di sicurezza
Siemens AG
3.3

Settings

After assembly and commissioning in line with the device manual, the
following parameter settings should be made for the safety function:
Safety parameters
Please enter following parameter via ULTRAMAT 6 menu:
Function
Number
Reference
Instruction Manual ULTRAMAT 6E/F, OXYMAT 6E/F
Protection against configuration changes
After configuration, the ULTRAMAT 6 access codes (function number 79) shall
be changed so that the device is protected against unwanted and
unauthorized changes/operation.
Checking the safety function after installation
After installation of the ULTRAMAT 6 a safety function test has to be carried
out (see chapter 4 "Maintenance" and 5 "Operation" of the instruction manual
ULTRAMAT 6E/F, OXYMAT 6E/F).
Using zero gas, i.e. N
For the test of the safety function it is fundamental to use a second reference
gas with a defined proportion of the component to be measured. The results
of the measurement must be within a range of ±5% (full span) of the
expected result.
SIL Safety Manual ULTRAMAT 6 Gas analyzers
A5E00729183-02
Function
70
Analog Output
, 4 mA must be measured at the analog output.
2
Select NAMUR/ 4-20 mA
10 of 19