VI. CORRECTIONS TO VOLUMETRIC FLOW
Changes in altitude and temperature can cause changes in air density,
thus affecting the output of the CAL-101. The output of the CAL-101 was
calibrated to STP (standard temperature and pressure) of sea level and
20°C. To compensate for differences in altitude or temperature from STP,
the following tables should be utilized:
Note: Altitude and Temperature are cumulative, i.e., you must correct
for both factors to obtain the most accurate calibration.
Altitude Correction Table
Altitude
0
1000
2000
3000
4000
5000
6000
Example:
At 3200 feet with a temperature of 29°C. The corrected volumetric flow
rate is
.5 LPM x 1.116 x 1.034 = .577 LPM (volumetric)
Corrected Out-
put (multiply by)
1.000
1.038
1.076
1.116
1.158
1.202
1.248
Temperature Correction Table
Temperature
Corrected (multiply by)
(°C)
-30
-20
-10
0
10
20
30
40
50
Page 9
0.829
0.863
0.898
0.932
0.966
1.000
1.034
1.068
1.102