Argus Optical Light Source Manuale d'uso - Pagina 11

Sfoglia online o scarica il pdf Manuale d'uso per Apparecchiature di prova Argus Optical Light Source. Argus Optical Light Source 14. Dual-wavelength laser source

5 Operation
5.8 Wavelength detection with Twin Mode and ARGUS SFP-OPM
Wavelength detection can be performed with the ARGUS OPM-SFP using the twin mode.
The bit sequence transmitted in twin mode is recognized in the ARGUS 300 with the OPM-
SFP plugged in and assigned to a wavelength.
To test the wavelength detection, the power measurement is started with an ARGUS OPM-
SFP on the ARGUS 300. Lightsource and ARGUS OPM-SFP must then be connected with
an optical fiber. It must be noted that only SC/APC connectors (Fig.) may be plugged into
the Lightsource connectors. SC/APC or SC/UPC can be plugged into the OPM-SFP.
SC/APC = GREEN
SC/UPC = BLUE
Due to different grinds of the fiber, only SC/APC and SC/APC as well as SC/UPC and SC/
UPC may be connected.
If the Lightsource and OPM have been connected, the twin mode will be started in the
plugged channel. After a few seconds, the ARGUS 300 display will change to a table view
showing the four possible wavelengths.
When using the twin mode in channel 1, the three wavelengths are detected one after the
other and the loss on the used transmission path is displayed.
11