aRIEquipment GAIA E Manuale - Pagina 10
Sfoglia online o scarica il pdf Manuale per Apparecchiature mediche aRIEquipment GAIA E. aRIEquipment GAIA E 11. Powered stretcher
In the process of use, the operator must follow the instructions, or the company will
not be liable for any risk.
2. When the stretcher is getting off the ambulance, it should be carried out on a flat
ground. To ensure safety, the car body must wait until the legs of the wheels are
fully extended in the air before the wheels can leave the car, so as to avoid the
ambulance falling forward and injuring the patients when the stretcher lands.
3. Illegal and barbarous operation is strictly prohibited, and overload operation is
not allowed to avoid damage caused by improper use of stretcher.
4. If the screws and nuts on the stretcher are loose, they should be tightened in time
and lubricated when necessary.
5. If the battery cannot be normally used, contact the manufacturer for treatment. If
the parts are replaced, the replaced parts shall be handled according to relevant
national regulations.
7. Maintenance
1. Always keep the stretcher clean and tidy.
2. Cleaning of the stretcher:
The cleaner used on the surface of the electric stretcher and the restraining belt is in
accordance with the neutral cleaning agent specified in 9.3 of the Hospital
Disinfection Supply Center Part 1: Management Regulations: pH 6.5-7.5, neutral
cleaner which is non-corrosive to metal.
When the hospital electric stretcher is used for the first time or after long-term
storage, it is necessary to clean the whole stretcher. You can use a neutral detergent
to clean the whole structure (remove the battery, pay attention to the circuit control
part should not touch the water during the whole cleaning process) Cleaning, the
main contact part of the product which we are using(mattress, guardrail, handle,
restraint belt) should be wiped and cleaned with a medical enzyme-containing
neutral detergent. After wiping with the cleaner, rinse with clean water and air dry.
Use 75% alcohol to wipe the main contact parts (handle, restraint band) before each
use.
After each use, the main contact part (handle, restraint belt) is wiped and cleaned
with a medical enzyme-containing neutral detergent. After the cleaner is wiped off,
rinse with clean water and air dry.
Clean the electric stretcher once a month, and wipe the cleaner every time you
clean it. After cleaning the water will evaporate.
Cleaning precautions:
a) Always remove the battery! Do not clean the electric stretcher when the battery
8