INTRODUCTION

This guide describes the steps for calibrating the ALCOSCAN breath alcohol tester
using the ALCOSIM breath alcohol simulator and an alcohol reference solution.
NOTE: For information about the general usage of the ALCOSCAN tester and
ALCOSIM simulator, refer to their respective instruction manuals.

ASSISTANCE

For assistance, contact your local authorized service provider or visit acs-corp.com.
WHEN TO CALIBRATE THE ALCOSCAN
TESTER
For optimal performance, calibration is required every 12 months or 5,000 tests. The
WARN result LED will blink blue on startup to indicate that the device needs to be
calibrated within the next 10 days or 100 tests. The WARN result LED will continue to
blink blue after startup if the device is overdue for a calibration.
WARNING: Removing the batteries for a long period of time may result in loss of date/
time and log events, causing a premature request for calibration.