Alcolizer Druglizer LE5 Manuale d'uso - Pagina 10
Sfoglia online o scarica il pdf Manuale d'uso per Unità di controllo Alcolizer Druglizer LE5. Alcolizer Druglizer LE5 16. Charging cradle
Anche per Alcolizer Druglizer LE5: Manuale d'uso (20 pagine), Manuale d'uso (19 pagine), Istruzioni per il montaggio (3 pagine)
![Alcolizer Druglizer LE5 Manuale d'uso](/uploads/products/server2/694511/694511.jpg)
3
4
5
6
REF: Druglizer LE5 Drug Tester User Manual V1
Technology
• Approximately 500uL (half a millilitre) of
saliva is required per drug test.
• If insufficient saliva is taken, an 'Invalid'
test result may occur.
• A Collector Indicator is located opposite
the Collector pad and will change to a blue
colour when sufficient saliva is taken.
Firstly, instruct the subject to insert the
collector pad into their mouth then:
• rub on inside of both cheeks;
• collect saliva around mouth and
under the tongue.
Ensure the sponge is not chewed.
• Instruct the subject to insert the Collector
into their mouth and rub the pad onto the
inside of both cheeks as well as under the
tongue and around the mouth.
• IMPORTANT: Druglizer LE5 training MUST
be undertaken in order to ascertain correct
saliva collection. Collect saliva until the
Collector indicator changes to a BLUE that
is consistent from edge to edge.
• When enough saliva is taken, the Collector
Indicator will change to a BLUE that is
consistent from edge to edge
Page 10
©
Alcolizer