Alcolizer Druglizer LE5 Manuale d'uso - Pagina 14

Sfoglia online o scarica il pdf Manuale d'uso per Unità di controllo Alcolizer Druglizer LE5. Alcolizer Druglizer LE5 16. Charging cradle
Anche per Alcolizer Druglizer LE5: Manuale d'uso (20 pagine), Manuale d'uso (19 pagine), Istruzioni per il montaggio (3 pagine)

Alcolizer Druglizer LE5 Manuale d'uso
The instrument is factory-calibrated. The instrument must only be repaired or calibrated by
authorized, qualified personnel.
The instrument may be damaged by liquid ingress. Please ensure that water and other liquids
are prevented from penetrating the instrument. If any fluid enters the internal parts of the
instrument, switch off the instrument immediately and return to authorised service centre.

12.2 Replacing sample module

Rather than returning the instrument to Alcolizer for service, the module can be removed and
sent in. This saves money and if the owner details are known and maintained with Alcolizer
Technology, arrangements can be made to forward a replacement sample module prior to
the calibration being overdue.
Remove and replace the sample module as follows:
a.
Undo the single screw on the back of the main module.
b.
Slide off the out-of-calibration sample module.
c.
Slide on the new sample module.
d.
Do up the single screw on the back of the main module.
Do not over tighten the sample module securing screw as this will damage the threaded portion.
Tighten only enough to prevent the sample module from sliding away from the main module.
a.
Following the replacement of the sample module, send the out-of-cal module back to
Alcolizer.
The instrument can be returned to Alcolizer Technology for calibration should the owner
prefer prior to the number of days warning message period expiring. To arrange calibration,
contact Alcolizer Technology.
Phone: 1300 789 908;
Web:
www.alcolizer.com
REF: Druglizer LE5 Drug Tester User Manual V1
Technology
CAUTION
c.
b.
Page 14
d.
©
Alcolizer