Direct Healthcare Group 56-310-B/1 Manuale d'uso - Pagina 6
Sfoglia online o scarica il pdf Manuale d'uso per Aiuto alla mobilità Direct Healthcare Group 56-310-B/1. Direct Healthcare Group 56-310-B/1 14.
11. Maintenance
Recommended at least once per year
Check that
General
The Walker feels firm/free from play.
The Walker does not rattle during manoeuvres.
The Walker is level and all of its castors are in contact with the
floor.
The Walker is not dirty
Armrest
Armrests are intact and clean.
Armrest widening is functional.
Handle
Grips are not damaged/dirty.
Handle adjustment is functional.
Battery
Check that the charging cable is connected to the control box.
Check that the remote control is connected to the control box.
Check that connections are made according to the wiring
diagram.
Check that the battery, actuators and control box are not loose.
Height adjustment
Raising and lowering are functional.
The Walker feels stable at maximum settings.
There is no play between the vertical frame and bottom frame.
Height adjustment clamps lock.
The electric actuator fitting must be free from play.
Height adjustment via remote control is
functional. The electric actuator must run
smoothly at a constant speed.
Frame parts
There is no mechanical damage.
There are no scratches.
End plugs/lower frame fitted
Castors/brakes
Castors roll easily/tread not damaged.
The castors are firmly fastened to the lower frame.
Castor brakes functional on all castors.
Handbrake functional.
6
BURE STAN D AR D/S /L OW /L O W S
Action
Clean with lukewarm soap solution or alcohol-based cleaning
agent (no petroleum products).
Clean; replace if damaged.
Fit lock screws/plates as required
Clean with lukewarm soap solution (no petroleum products).
Replace handle grips.
Fit lock screws/plates as required
Fit new charging cable; this must always be connected to the
control box (IP65 protection).
Connect or fit new remote control as required.
Refer to the wiring diagram in the Technical manual.
Tighten; replace with new fasteners as required.
Fit new guide sleeves (in frame) or end plugs (in chrome tubes) as
required.
Tighten all lock bolts.
Replace clamps.
Tighten the attachment fitting concerned; replace bolts/lock nuts
as necessary.
Check that the battery/remote control/ actuator are connected in
accordance with the wiring diagram. Charge the battery. Contact
customer services.
If damage is present, contact GATE's customer services.
Touch up as necessary.
Fit new end plugs
Clean or replace castors. NOTE: Castors are always fixed to the
lower frame with thread locker or lock
nuts. Do not dismantle the castors; replace the whole castor
assembly instead.
Tighten the castor bolt and fit locking nut or use thread locker
(depends on model).
Replace castor assembly.
Adjust the brake or fit new castors.