Alcovisor Jupiter Manuale - Pagina 5
Sfoglia online o scarica il pdf Manuale per Strumenti di misura Alcovisor Jupiter. Alcovisor Jupiter 8.
Anche per Alcovisor Jupiter: Manuale di istruzioni (12 pagine), Manuale di istruzioni (12 pagine)
NOTE: When using dry gas (PAS recommends 0.080% dry gas
concentration) for calibration, the alcohol concentration printed on the label
of the dry gas cylinder must be corrected to account for altitude when
conducting tests at or above an altitude of 250 feet above sea-level. Please
refer to the Operating Manual for specific instructions for altitude
corrections.
CALIBRATION: (Using an approved Wet Bath Simulator or Certified
Dry Gas with proper Regulator)
1. (It is recommended to set your AUTO OFF timer to 10 minutes in order to complete Calibration) Select the Settings
and select "Cal.".
Icon
2. Enter the password 119119.
3. Select Dry Gas Calibration
4. Enter the concentration of the calibration solution/cylinder. Accuracy should be to one decimal place. The preset
unit is mg/100mL. For example, enter 80.0 for 0.080% or enter 100.0 for a 0.100% solution.
Note: Remember to use the
elevation.
More examples are in the following table:
Mg/100mL
20.0
50.0
80.0
100.0
150.0
5. Select
to confirm.
High-Altitude Correction Chart
‰ & g/L
%
0.02
0.20
0.05
0.50
0.080
0.80
0.10
1.00
0.15
1.50
to adjust for altitude if needed based on your