8.11.4 The hooks
Hook attachment
1
Rubber suspension
2
Hook
3
8.110.5 Air inlet/power outlet
Air outlet
1
Power inlet
2
Fuse holder
3
ON/OFF
4
8.12 Filter Maintenance
The Control Unit has a light that when illuminated indicates a Service is required.
1. Change the air filter: Remove air filter holder by pressing the
plastic catch downwards and lifting the air filter holder up.
Take out the filter and put in a new one and put the filter
holder back.
8.13 Fuse
2. Check the fuse: Check the fuse and if required add a spare in
the holder for the fuse in the fuse holder.
8.14 Rotor Valve Maintenance
The Control Unit has a light that when illuminated indicates a
Service is required. Open the housing by remove the 4 screws from
the bottom part.
3. Grease the rotor valve: Press down the cap of the rotor valve
release the lock pin. Lift of the cap and put a thin layer of grease
(Silicone grease dow Corning Molykote PG21, 613000103) on
the bottom side of the cap. Assemble back again.