Direct Nexus MD758035ND-01 Manuale d'uso - Pagina 18

Sfoglia online o scarica il pdf Manuale d'uso per Prodotto respiratorio Direct Nexus MD758035ND-01. Direct Nexus MD758035ND-01 20. Compressor nebulizer

11. Troubleshooting Manual
Sign
Excessive Sound
Water drops
in the air pipe
Note: Please, contact with your supplier for all problems.
D str butor:
Probable Cause
No filter in the unit
Main Compressor button
is turned off
Nebulizer has not been
cleaned after previous use
Air pipe bent.
Filter blocked
No medicine available
Cover not properly closed.
Excessively filled
Nebulizer has been
washed but not dried.
Solution
Put a filter into the unit
Turn on the unit
Clean the unit
Remove bending and
entanglement in air pipe
Remove the filter.
Add sufficient amount of
medicine specified by the
physician to the container.
Adjust the cover again.
Add appropriate amount of
medicine to the container,
connect the air pipe to the
air compressor and open
the unit.
Turn on and off the
Nebulizer with your finger
over and over again until
the drops disappear.
Producing Company:
Medisec Tıbbi Cihaz ve Ürünleri
İthalat İhracat San. ve Tic. Ltd. Şti.
Başpınar (Organize) OSB Mahallesi O.S.B. 2. Bölge
Celal Doğan Blv. No:40 Şehitkâmil/Gaziantep/Türkiye
Tel: +90 342 337 88 75 Fax: +90 342 337 88 76
www.medisec.com.tr
18