Calex PyroUSB PUA2 Manuale dell'operatore - Pagina 12

Sfoglia online o scarica il pdf Manuale dell'operatore per Termometro Calex PyroUSB PUA2. Calex PyroUSB PUA2 17.

Address
Length (words)
0x11
1
0x12
1
0x13
1
0x14
1
0x15
1
0x17
1
0x18
1
Notes:
1
All temperature are in tenths of degrees C
2
With the exception of addresses 0x17 and 0x18, all write operations are saved to non-
volatile memory
3
For further information please refer to http://www.modbus.org/specs.php
4
The default Modbus slave address of the sensor is 1; it is also possible to use slave
address 255 to communicate. 1 sensor may be connected per COM port.
12
Description
Maximum Temperature
Filtered Temperature
Unfiltered Temperature
Sensor Temperature
Status (bits active high):
Bit 0: Measurement error
Bit 1: Sensor temperature low
Bit 2: Sensor temperature high
Bit 3: Object temperature low
Bit 4: Object temperature high
Reflected Temperature
Not saved to non-volatile memory
Use this register when writing frequently
Emissivity (1 LSB = 0.0001)
Minimum 0.1000, Maximum 1.0000
Not saved to non-volatile memory
Use this register when writing frequently
R/W
R
R
R
R
R
R/W
R/W