Xerox 6180N - Phaser Color Laser Printer Manuale introduttivo - Pagina 28

Sfoglia online o scarica il pdf Manuale introduttivo per Software Xerox 6180N - Phaser Color Laser Printer. Xerox 6180N - Phaser Color Laser Printer 34. Color laser printer
Anche per Xerox 6180N - Phaser Color Laser Printer: Manuale di installazione (34 pagine), Scheda tecnica (2 pagine), Manuale d'uso (32 pagine), Manuale d'uso (28 pagine), Manuale di installazione (2 pagine), Manuale di sicurezza (38 pagine), Nota di rilascio (2 pagine), Manuale di installazione (2 pagine), Manuale introduttivo (37 pagine), Manuale (34 pagine), Manuale (34 pagine), Manuale d'uso (32 pagine), Manuale di installazione e funzionamento (8 pagine), Manuale del valutatore (16 pagine), Specifiche tecniche (4 pagine)

Xerox 6180N - Phaser Color Laser Printer Manuale introduttivo

Reset the Xrite Spectrophotometer

24
Changing the Retain PDL Setting
Perform the following steps to modify the number of PDLs that can
be retained on the DocuSP.
1. On the DocuSP interface, select Setup and System
Preferences. Select the Job Policy tab.
2. Ensure that the setting for Retain PDL files of last 50 jobs is set
as enabled.
3. Select [OK]. Do not restart the system if you are prompted to do
so.
4. In the background, right-click to open the menu. Open a
terminal window.
5. Login as root user.
6. Type: /opt/XRXnps/bin/setretainPDL # and press <Enter>,
where # is a number between 1 and 100. A value of 0
maintains the current setting.
7. Close the terminal window. Restart the DocuSP software.
Perform the following to reset the spectrophotometer:
Ensure that the calibration strip that matches the serial number on
the spectrophotometer is available. If not, order a new one before
continuing.
1. On the DocuSP controller background menu, open a terminal
window.
2. At the % prompt, type:
cd /opt/XRXnps/XRXcalLib/bin
3. From the % prompt, type:
. /resetDTP
This will restore the spectrophotometer to the factory default
settings.
After the reset is complete, follow the standard procedure to
calibrate the spectrophotometer using the Xrite calibration strip.