Maintenance

NOTE: Inspect suction tubing and collection for leaks, cracks, etc. before each use.
NOTE: Do not attempt to open or remove the suction cabinet. If a service is
required then return your suction to an authorized service center. Opening or
tampering with the unit will void warranty.
Bacteria Filter Change:
• B acteria filter should be replaced as needed. If overflow occurs, change the filter
immediately.
• R emove filter by disconnecting it from tubing connected to pressure adjusting
valve and tubing connected to collection lid assembly.
• R eplace with a new bacteria filter and connect to above tubing. Additional filter
can be purchased from an authorized dealer.
• D o not substitute any other material for this bacteria filter. Improper substitution
may lead to contamination or poor performance.
Fuse Change:
• Use only the same type of fuse to replace the broken fuse.
• F use specification: Miniature Glass Fuse 5 x 20mm, Fast Blow (Quick Acting),
F2A/250V.
NOTE: When this unit or accessories reach the end of their useful lives
(depends on using conditions), residues in collection bottle, tubing and filter may
cause respiratory infection. Dump the waste collection bottle, tubing and filter into
a specific waste bin labeled with medical waste.