Anche per Direct Supply Panacea: Manuale d'uso (16 pagine), Manuale d'uso (5 pagine), Manuale d'uso (16 pagine), Manuale d'uso (2 pagine), Manuale d'uso (5 pagine), Manuale d'uso (2 pagine), Manuale d'uso (2 pagine), Manuale d'uso (2 pagine), Manuale d'uso (2 pagine), Manuale d'uso (2 pagine), Manuale d'uso (16 pagine)

NOTE: Before using the Mattress System, please remove current mattress,
mattress replacement system or overlay systems from the bed.
• Mattress set up.
Make sure the mattress fits snugly on the bed frame without any gaps on the side.
• Mattress Replacement: Unroll the MotionTurn mattress replacement and place it
directly on the bed frame and attach it firmly with the straps provided. Hang the
MotionTurn control unit on the footboard, make sure the mattress bolsters and/or
air pad plugs are connected properly and check that the mattress hose assembly is
connected securely to the control unit.
• Plug the hospital grade power cord provided with the unit into a three-pronged hospital
socket. The amber light will illuminate. Press POWER key (SB) to green light position
and then press the Select key (S) until MAX Inflate (W) mode is activated.
The compressor will come on and inflate the mattress.
• Using the SOFT / FIRM keys (K), set resident comfort pressure settings.
7-segment display will show resident comfort settings from 0 (lowest pressure)
to 9 (highest pressure).
• To set STATIC (T) mode, press Mode key (M) until the "THERAPY" LED (T) lights up.
• To set ROTATION therapy mode, press the Mode key (M) until the desired Rotation
function is selected and corresponding LED lights up. Adjust Rotation therapy
settings using the Rotation angle (RA) and Rotation cycle time (TT) keys.
• To activate Low Air Loss (Y) press the Select key (S) until Low Air Loss LED lights up
• To set ALARM SILENCE, press the ALARM SILENCE key (AS) until the indicator
light goes on.
• To set the LOCK OUT Function, press the LOCK OUT key (LO) until indicator light
goes on.
• The MotionTurn mattress system is equipped with Fowler function (U) which when
pressed in the resident's Fowler position of 35 degrees or greater will inflate the
mattress to higher pressures to eliminate resident bottoming. During Fowler mode,
Rotation mode is disabled. When Fowler mode is activated the Rotation function
will cease if activated and Fowler mode will be activated.
• For CPR or quick deflation, disconnect mattress hose connector from the control
unit by simply pressing the latches on the quick disconnect connectors and pulling
the connector coupling from the control unit in a straight outward direction. If the
mattress is equipped with bolsters and/or air pad, unplug the corresponding deflate
plug located at the foot end of the mattress.