Preventive Maintenance & Storage
Control Unit
The controller contains no user serviceable parts. Any service required on the
controller should be referred to your Direct Supply representative.
Mattress
The Panacea Air Pro Mattress requires routine cleaning at least twice a week and
maintenance to ensure top performance. The caregiver should familiarize themselves
with the following information before maintaining the mattress.
Semi-Permeable Top Cover
The Top Cover must be in place when the resident is on the mattress and should be
removed and laundered periodically for sanitary reasons. Laundering instructions are
provided below. Additional Top Covers are available for laundry rotation. Please contact
your Direct Supply account manager for further information on additional Top Covers.
The Top Cover should be wiped off with an EPA registered disinfectant solution, rinsed
with fresh water, and wiped dry to remove stains and body fluids between launderings.
Caregivers should take care to make sure the Top Cover is not damaged by abuse.
Protect the Top Cover and air cells from cigarette burns, sharp objects, tears from
sharp objects and pin holes. DO NOT CLIP OR PIN ITEMS TO THE TOP COVER SUCH
AS NURSE CALL CORDS, RESIDENT LINES OR OTHER CABLES.
The advanced microclimate top cover is a performance fabric with a polyurethane
barrier on the top side and a stretchable membrane on the underside. The polyurethane
coating allows air and water vapor to pass through while stopping liquids from
penetrating the cover. The cover has a slick and soft surface to minimize shear forces
on the skin. The material is very soft to permit easy conformity to the skin surface.
Laundering the Semi-Permeable Top Cover
The Top Cover should be disinfected prior to washing with an EPA registered
disinfectant, washed with a commercial soap in a washing machine with a water
temperature of less than 150°F (65°C) with a full rinse cycle. The covers should be
immediately placed in a dryer set to low/delicate heat (less than 150°F/65°C) or hung
to air dry. Once dry, the covers can be bagged in plastic or immediately returned for
resident use.
The Air Cells
Air cells can be replaced easily in the event they become damaged. Immediately
replace any damaged air cells as they may prevent the mattress from reaching full
inflation. Air cells can be replaced by unsnapping each end of the air cell from the
mattress base and detaching the air inlet from the manifold.
Storage
1. Always store the support surface flat on a clean, level surface.
2. Avoid storage of other equipment on top of the support surface.