FRAP on the Fly Calibration
Before you can use FRAP the laser must be calibrated
7
• Find an empty piece of glass without cells/fluorescence
• Set the exposure time around 100 - 200ms
• Open ILAS2 with icon
• Open calibration tab and put the power of selected laser to 100% (in ILAS2 window)
• Activate the laser [G] and press " Show Live"
• Place laser in upper left corner by moving the red cross, press [H]
• Repeat for lower right corner
o If the spot isn't visible or too big and bright change power and/or exposure time
o view is rotated, so move laser in other direction
o keep in mind to move red cross into left and right corners of the calibration screen
• Calibrate with button [i]
• To check if your calibration was successful open " on the fly" tab
• Change time to 3000
• Increase the power of your laser and press " Show Live"
• Click in the image and check if laser pulse in at the position of your mouse pointer
Troubleshooting - FRAP
FRAP calibration error
• Notification: Calibration canceled (exceeded 15 bits)
Possible errors
• Emitting fluorophore or auto fluorescence
o Ensure that the area of the slide really empty
• Too much laser power or exposure time
o Reduce laser power in the ILAS2 window and/or exposure time in Acquire window