Power Supply TN-PSE30W A (for regions with 230 V power only)
Brightness adjuster
Changes the illumination
brightness when rotated.
Power switch
Input rating
Voltage fluctuation
Output rating
Internal fuse
Operating
environment
Storage
environment
Protection class
Power cord
22
AC input connector
Power cord
Connect the specified power cord here.
230 VAC, 50/60Hz, 0.3 A
±10%
6 VAC, 5 A
Fast acting type, F1A/250 V, 5.2 × 20 mm
Altitude:
Temperature:
Relative humidity: 85% RH max.
Pollution level:
Installation:
Temperature:
Relative humidity: 90% RH max.
Class I
Use only the following power supply cord.
Using the wrong power cord could result in
danger or fire. The protection Class I equipment
should be connected to PE (protective earth)
terminal.
• For 220 to 240 VAC area
Approved according to EU/EN standards, 3
conductor grounding Type H05VV-F, 3 m
long maximum, rated at 250 VAC minimum.
Using Accessories
Input voltage display
Check that the voltage displayed matches the
voltage used in your area. If the voltages do
not match, do not use the power supply and
contact your nearest Nikon representative.
Output connector
Connect the
lamphouse cable here.
(Rear view)
2000 m max.
0 to +40°C
(indoor use only)
(no condensation)
Degree 2
Category II
-20 to +60°C
(no condensation)
IV