_TEST PROGRAM
@ TEST - 04 : Turning on the C. LED (Calibration LED) and M. LED (Main LED)
Press this
key to quit
TEST-04.
• Turning on the C. LED: Press the [CLED OFF] key to change the display into
[CLED ON], and the C. LED will light up. This mode is used to check the CHA and
CHB signals.
• Turning on the M. LED: Press the [MLED OFF] key to change the display into
[MLED ON], and the M. LED will light up cotinuously. This mode is used to adjust
the position of the M. LED.
(5) TEST - 06 : Displaying the CHA and CHB phase count offsets and the CHA - FIL
and CHB - FIL signal peak values
Press this
key to quit
TEST-06.
• Displaying the CHA and CHB phase count offsets
This test displays the digital read - out of the phase difference offset after A/D
conversion is made when the C. LED is on. Make sure that the following readings
are observed:
Offset: - 00200 to +00200 ;
Variation: 0010 or less
• Displaying the CHA - FIL and CHB - FIL signal peak values
This test displays the digital read - out of the peak values of the CHA - FIL and
CHB - FIL signals after A/D conversion is made when the C. LED is on.
CHA: 150 to 200 ; CHB: 150 to 200
Adjust VR7 CAL until the reading of the CHA - FIL and CHB - FIL peak values
falls between 150 to 200.
(it is advisable that this adjustment be carried out during TEST - 0D)
• 0V display: Make sure that the level measured is 000. If read - out is not 000, the
A/D converter is damaged.
C6 - 51
Repair Manual