BLADDERSCAN BVI 6100 Riferimento rapido - Pagina 2

Sfoglia online o scarica il pdf Riferimento rapido per Apparecchiature mediche BLADDERSCAN BVI 6100. BLADDERSCAN BVI 6100 2. Bladder volume

BLADDERSCAN BVI 6100 Riferimento rapido
8.
Press and release the
A scanning symbol
Scan
button
9.
Hold the probe steady until the scan is finished.
When you hear the end-scan tone, the scan is complete.
10.
Using the Results screen, verify that the aim was accurate.
NO ARROW
If no arrow is displayed, then the
bladder was completely contained
within the scanning cone. The
measurement is accurate.
11.
Note the patient's bladder volume.
The BVI 6100 instrument will not save this measurement after a new scan is begun.
BladderScan, the BladderScan symbol, Verathon, and the Verathon Torch symbol are
trademarks of Verathon Inc. © 2007, 2010, 2015 Verathon Inc. 0900-0589-10-60
Scan
button.
appears in the upper right-hand corner of the LCD screen during the scan.
If a solid arrow is displayed, then
the bladder was mostly inside the
scanning cone. The results are
satisfactory, but in order to ensure
accuracy, Verathon recommends
that you re-aim in the direction of
the arrow and perform the scan
again.
verathon.com
0123
https://my.scanpoint.com
SOLID ARROW
Corporate Headquarters:
Verathon Inc.
20001 North Creek Parkway
Bothell, WA 98011, USA
Tel: 800.331.2313 (USA & Canada only)
Tel: 425.867.1348 Fax: 425.883.2896
FLASHING ARROW
If a flashing arrow is displayed,
then the scan is off-target, and the
bladder was not enclosed within
the scanning cone. Re-aim the
probe in the direction indicated by
the flashing arrow and perform the
scan again.