BVI 9400 Calibration Quick Reference
BVI 9400 Calibration Quick Reference
www.verathon.com https://my.scanpoint.com
For more information, call 800.331.2313 (in the U.S. and Canada) or contact your local Verathon Medical
Verathon Corporate Headquarters: 20001 North Creek Parkway, Bothell, WA 98011, USA. Phone: +1.425.867.1348
VM Europe B.V.: Boerhaaveweg 1, 3401 MN IJsselstein, The Netherlands, Phone:+31.30.68.70.570
Calibrate instrument
From the Tools menu, select Calibrate Instrument. When
the Calibration screen opens, click the CALIBRATE button.
ScanPoint
then automatically scans and analyzes the
®
calibration target.
IMPORTANT! Do not remove the Probe from the calibration
chamber until four arrows appear on the Probe's LCD
screen. Do not disconnect the Charger and Wireless Hub
from the computer or otherwise break the connection
between the BVI 9400 and the ScanPoint
until all processing is complete.
Successful calibration
If ScanPoint
was able to locate the calibration target,
®
the instrument analyzes the scan data to ensure that it
meets the calibration parameters. If necessary, the
instrument automatically rescans the phantom. When
calibration is complete, a 'Calibration Successful' message
is displayed on your computer.
To print a record of the calibration, click: Click Here for
Printable Certificate of Calibration.
Failed calibration
If ScanPoint
was unable to locate the calibration target,
®
it displays an error message and instructs you to ensure
that the calibration chamber has enough water and that
the Probe is seated properly in the calibration system.
You may choose to calibrate again or to quit the
calibration process.
NOTE: If you are unable to calibrate, contact your local
Verathon Medical
representative or Verathon Medical
®
Customer Care at 1.800.331.2313.
End calibration
To terminate the calibration procedure and end
communication with the ScanPoint
CANCEL (X) button on the BVI 9400.
Remove the Probe from the calibration chamber lid and
dry it with a clean, soft cloth.
To order additional rolls of paper (0800-0319) or batteries (0400-0066), contact Customer Care at 800.331.2313.
host computer
®
®
, press the
®
Calibrate
representative.
®
0900-1447-01-60