ABBERFIELD CT22 Series Manuale tecnico e di servizio - Pagina 5
Sfoglia online o scarica il pdf Manuale tecnico e di servizio per Interruttore ABBERFIELD CT22 Series. ABBERFIELD CT22 Series 8. Coin time switch
ABBERFIELD INDUSTRIES - CT22 coin time switch TECHNICAL DETAILS
Coin
Diameter
Minimum
C1
C2
C3
C4
C5
C6
C7
C8
The normal calibration procedure is all that is
ever needed and the following information is
for special circumstances only.
Sometimes there is a desire to vary the standard
calibration criteria, to invalidate a closely matched
fraudulent coin or increase the certainty of correct
validation of closely matched coins. Understanding
the calibration criteria and recording set values in
a table will allow evaluation of the equipment's
operation and from this it can be seen where
adjustments to the calibration criteria will help in
achieving a particular result. First understand that
there are three measurements being taken for
each coin, (optical detection of diameter and
"induced" electronic readings referred to as
amplitude and period. For each of these parameters
there is a reference number that is set into the coin
validator calibration tables. It is possible to write up
a table indicating these readings.
1.
Follow steps 1. to 7. from the calibration
procedure and bring the desired coin on to
the screen.
2.
Press the SET button this will show the first
entry of the coin table, it should be between
3000 and 7500.
3.
Enter the displayed values for all coins in the
table provided, this will then help to show how
far each parameter can be adjusted before
overlapping another coin and possibly
causing miscredits. Also if any adjustments
made to parameters cause problems then
you have a record of the original values that
you can enter back into the validator.
4.
To move to the next coin parameter press the
NEXT button.
5.
After all parameters for that coin have been
displayed the screen will return to C1-C8.
6.
To advance to the next coin press the NEXT
button.
7.
To display parameters for the coin press the
SET button.
32 Cross St, Brookvale, Sydney NSW 2100, Australia
Adjusting Calibration
Adjusting Calibration
Period
Maximum
Minimum
Tel: (02) 99392844 Fax: (02) 99383462 Email: [email protected]
page 5
Amplitude
Maximum
Minimum
8.
Repeat steps 3 to 7 until all coins have been
recorded.
9.
An overlap in all 3 parameters (diameter,
period and amplitude) between 2 or more
coins will cause miscredits of coins which
fall into these parameters, but ONLY if they
occur simultaneously.
10. To change parameter press the SET button
while that parameter is on the screen. The
last digit (right side) will flash, this is the digit
that can be adjusted.
11. To advance the value of the flashing digit
press the NEXT button. The digit can then be
scrolled from 0 to 9 and back to 0.
12. To move to the next digit press the SET
button.
13. Repeat for all digits and once the first digit
(left side) is flashing then pressing the SET
button. The display will return to the coin
number, C1-C8.
14. Press the NEXT button to advance to the next
coin and repeat steps 10 - 13 to change other
coin parameters.
15. After adjusting the desired parameters make
sure the screen is showing a coin number C1-
C8. Remove power to the validator.
16. Set the dip switches to their normal operating
positions and re-apply power.
17. Test the validator to see if the adjustments
made do not cause problems with any of the
coins normally used.
NOTE: Some time a difficult fraudulent coin can be
"calibrated" so as to record its reference numbers
(Diameter Period and Amplitude). Then the
validator is recalibrated to remove the fraudulent
coin details, but the numbers taken can be used to
determine how best to adjust the Diameter Period
and Amplitude readings for the valid coin set.
Brochure # 1003-4
Maximum