Discovery Labs Warming Cradle 801-000 Manuale operativo - Pagina 5
Sfoglia online o scarica il pdf Manuale operativo per Apparecchiature di laboratorio Discovery Labs Warming Cradle 801-000. Discovery Labs Warming Cradle 801-000 20. Dry block heater
5
•
Audible signals that indicate:
– the unit is connected to the power supply
– the preset reservoir temperature of 44°C
– the vial has been warmed for 14 minutes
– the 15-minute warming cycle is complete
– removal of a vial prior to the completion of the warming cycle
– immediate service is required
•
LCD and multicolor LEDs that provide information regarding status,
temperature, and time
•
Temperature regulation to the preset temperature of 44°C
WARMING CRADLE
®
uses to maintain the preset temperature of 44°C
use at all times.
Specifications
•
Input voltage is 110 V AC, 50/60 Hz
•
Input current is 0.75 A at 110 V AC
•
Two 40 W heating elements are used for warming the vials
•
The temperature is controlled to
•
If the warming temperature is outside of the specified range, an audible
signal indicates that the unit requires immediate service
•
Classification according to UL61010-1 with regard to electrical safety:
– For indoor use only
– For use at altitudes up to 2000 meters
– Ambient temperature range is 5°C to 40°C
– Maximum relative humidity 80% for temperatures up to 31°C
– Humidity decreasing linearly to 50% relative humidity at 40°C
– Input voltage fluctuations within
– Transient Overvoltage Category II
– Pollution Degree 2
±
1°C has been reached
is designed to remain plugged in between
±
±
1°C
±
10% of the nominal input voltage
±
1°C
1°C and be ready to