Abbott StarClose SE Manuale - Pagina 13

Sfoglia online o scarica il pdf Manuale per Apparecchiature mediche Abbott StarClose SE. Abbott StarClose SE 15. Vascular closure system deployment steps

Abbott StarClose SE Manuale
StarClose SE
Vascular Closure System
Use of Safety Release
If excessive resistance is felt while advancing the Thumb Advancer, pull back the Thumb Advancer.
Slide the Safety Release forward to collapse the Locator Wings.
Check that the number "2" on the Plunger exits the number window completely.
Gently push down on the patient's body with the left hand. Remove the Clip Applier with the right hand without rocking or twisting.
©2011 Abbott. All rights reserved. ANZ 080091 10/11