Note: If the deviation between the actual
temperature and the calibration temperature is
within ±2℃ during calibration, a single point
calibration can be performed. If the deviation
between the actual temperature and the calibration
temperature exceeds ±2 ℃, a three-point calibration
is required.
5.5 Restore the factory settings
1. Switch power ON the instrument, Press Temp
and Prog buttons at the same time to enter into
calibration mode while the instrument is in self-
check mode. (Please note that the instrument cannot
enter into calibration mode if the self-check is
complete.)
2. After entering the calibration mode, press both
the Prog and Time keys to reset, the interface will
display "RESET Success," and the instrument will
restart automatically.

6 Trial Run

Make sure the required operating voltage
and power supply voltage match.
Ensure the socket must be properly grounded.
Add the medium into the vessel
Place vessel on the block.
Plug in the power cable,ensure the power is
on and begin initializing.
Set the target parameters or pragraming
Start working
Observe the LCD display
Stop working, and power off.
If these operations above are normal, the device is
ready to operate. If not, the device may be damaged
during transportation, please contact
manufacturer/supplier for technical support