dlb AIJGO-61 Manuale di istruzioni - Pagina 6

Sfoglia online o scarica il pdf Manuale di istruzioni per Strumenti di misura dlb AIJGO-61. dlb AIJGO-61 10. Resistance meter

dlb AIJGO-61 Manuale di istruzioni

Turning on

Turn on the device with the "On/Off" button. After switching on, a text similar to the
following appears on the screen containing the device's data:
This is visible for 10 seconds. If you do not want to wait this time, but want to move
on faster, press any button except "On/Off". You will then see a screenshot similar to
the one below:
Measuring voltage
Temperature
Measurement
The AIJGO-61 is suitable for measuring a wide resistance range, between 2 Ω and
1 TΩ. This also enables the measurement of equipment grounding resistance and
the resistance of ESD protective worksurfaces to ground. Thus there is no need
to use other resistance meter during ESD qualification or compliance verification
measurements.
Press and hold the "Measure" button until the measurement result is displayed. This
will remain visible until you press the "Measure" button again.
...MEASURING...
24.5°C
When measuring temperature and humidity, make sure that you do not cover the
sensor on the bottom of the device with your hands or other objects, and do not
place the device on a cooled or heated surface, as this may affect the measurement
results.

Storing measurement results

The measurement results can be stored by pressing and holding the "MODE" button,
after this a message is shown on the display for 3 seconds. If any blank space is still
available in the measurement list, a message similar to the following will be displayed:
AIJGO-61
RESISTANCE METER
S/N: 1234
Ver: v1.02
_____
10V
24.5°C
100V
42.1%RH
42.1%RH
187 MΩ
100V
24.5°C
42.1%RH
6
Place to display the
resistance value
Relative Humidity